The data were obtained from two Interlaboratory comparisons (ILC1 and ILC2) involving six Spanish biodosimetry laboratories dedicated to harmonize and establish a standardized national protocol. Both ILCs were organized by La Fe Hospital in Valencia. Two established techniques were applied among the six participant laboratories: the dicentric chromosome assay (DCA) and the cytokinesis-block micronucleus (CBMN) assay. In the first ILC, blood samples were irradiated at 2.9 Gy (sample A) and 1.4 Gy (sample B). In the second ILC, a 2.63 Gy exposure was delivered, generating a partial-body sample (sample A) and a whole-body sample (sample B), the latter prepared by mixing equal proportions of irradiated and non-irradiated blood. The data include the raw values to obtain the calibration curves using the software Biodose Tools v 3.7.2 for each laboratory, and the aberration distribution for the two evaluated samples (sample A and sample B) for each comparison.
Biodose Tools, 3.7.2
METHODOLOGICAL INFORMATION
-
Description of methods used for collection-generation of data
Blood samples from one healthy donor were obtained by venipuncture, collected into heparinized tubes, and irradiated ex vivo. For each ILC, two heparinized tubes containing 3 mL of blood were labelled as Sample A and Sample B and sended to the participant laboratories. Each laboratory cultured lymphocytes from the blood they received following their own protocols and the IAEA and ISO standard recommendations. Two laboratories used a semiautomated method to capture the metaphases that were then analysed manually (counting aberrations), and the remaining laboratories performed the cytogenetic analysis directly under the microscope.
-
Methods for processing the data:
Biodose Tools v3.7.2
Can be found freely online at https://www.reneb.net/software/
-
Instrument- or software- specific information needed to interpret the data:
OpenOffice
Biodose Tools v3.7.2
Can be found freely online at https://www.reneb.net/software/
-
Instruments, calibration and standards information:
All samples were irradiated ex vivo in a methacrylate tank with water at 37°C using the TrueBeam HDR accelerator located at the Radiation Oncology and Medical Physics Service of La Fe Hospital (Valencia, Spain). Irradiations were performed using a 6 MV photon beam (6X) with a dose rate of 600 MU/min. The irradiation field at isocentre was 8 × 16 cm², and the samples were positioned at a depth of 9 cm. The source-to-surface distance (SSD) was set to 100 cm minus the depth, resulting in an effective SSD of 91 cm. Exposures were delivered using parallel-opposed fields, with the gantry positioned at 90° and 270°, corresponding to left-inferior (LI) and left-dorsal (LD) orientations. The irradiation was administered at the 100% isodose level.