GEOMAR moorings are typically equipped with instruments recording pressure, temperature, conductivity, dissolved oxygen and current velocity. Instruments with pressure, temperature, conductivity and oxygen sensors were calibrated in situ immediately prior to and after a mooring deployment period by attaching them to the CTD frame during CTDO casts. Correction terms were then developed from the difference between the sensor readings and the calibrated CTDO data during several minute long calibration stops. These correction terms were then applied to the full deployment periods. This ensured best data quality with recognition of potential sensor drifts and also allowed for the estimation of calibration and measurement errors (Hahn et al. 2014, Bittig et al. 2018, Berx et al. 2019).
instrument = Label:kpo_1210_02 File:kpo_1210_instr_02.nc Type:Teledyne RDI QuarterMaster 150 SN:14910instrument = Label:kpo_1210_03 File:kpo_1210_instr_03.nc Type:GEOMAR TD-Logger SN:55 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1210_04 File:kpo_1210_instr_04.nc Type:Teledyne RDI LongRanger 75 SN:12530instrument = Label:kpo_1210_05 File:kpo_1210_instr_05.nc Type:Seabird MicroCat 37-SM mit P SN:10709instrument = Label:kpo_1210_06 File:kpo_1210_instr_06.nc Type:Aanderaa Optode 4330 SN:00379 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1210_08 File:kpo_1210_instr_08.nc Type:Seabird MicroCat 37-SM mit P SN:10708instrument = Label:kpo_1210_09 File:kpo_1210_instr_09.nc Type:Aanderaa Optode 4330 SN:00375 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1210_10 File:kpo_1210_instr_10.nc Type:Seabird MicroCat 37-SM SN:02614 depth/pressure data was at least partially derived and not directly measured, consult quality flags for detailsinstrument = Label:kpo_1210_12 File:kpo_1210_instr_12.nc Type:Nortek Aquadopp Seriell SN:P26209-37instrument = Label:kpo_1210_14 File:kpo_1210_instr_14.nc Type:Nortek Aquadopp Seriell SN:P26209-30instrument = Label:kpo_1210_15 File:kpo_1210_instr_15.nc Type:Nortek Aquadopp Induktiv SN:P26209-09instrument = Label:kpo_1210_16 File:kpo_1210_instr_16.nc Type:Seabird MicroCat 37-SM SN:02618 depth/pressure data was at least partially derived and not directly measured, consult quality flags for details